Evaluate Continuing Calibration Report Data Path : P:\HPCHEM1\FID_A\Data\FA010318\ Data File : FA001994.D Signal(s) : FID1A.CH Acq On : 3 Jan 2018 13:49 Operator : UA/AJ Sample : 200 PPMV CCC Misc : ALS Vial : 20 Sample Multiplier: 1 Integration File: autoint1.e Quant Time: Jan 03 16:06:57 2018 Quant Method : P:\HPCHEM1\FID_A\GASES\Method\FA122717 GASES.M Quant Title : QLast Update : Wed Dec 27 17:07:17 2017 Response via : Initial Calibration Integrator: ChemStation Volume Inj. : Signal Phase : RT-U Plot Signal Info : 30M x 0.32mm Min. RRF : 0.000 Min. Rel. Area : 50% Max. R.T. Dev 0.50min Max. RRF Dev : 25% Max. Rel. Area : 150% Compound AvgRF CCRF %Dev Area% Dev(Min) ------------------------------------------------------------------------- 1 Ethane 64.476 60.822 5.7 92 -0.24 2 Ethylene 61.983 58.248 6.0 90 -0.18 3 Methane 40.505 40.869 -0.9 105 -0.02 Evaluate Continuing Calibration Report - Not Founds -------------------------------------------------------------------------- (#) = Out of Range SPCC's out = 0 CCC's out = 0 FA122717 GASES.M Tue Jan 09 04:34:11 2018