Evaluate Continuing Calibration Report Data Path : P:\HPCHEM1\FID_A\Data\FA112917\ Data File : FA001891.D Signal(s) : FID1A.CH Acq On : 29 Nov 2017 13:08 Operator : UA/AJ Sample : 200 PPMV CCC Misc : ALS Vial : 1 Sample Multiplier: 1 Integration File: autoint1.e Quant Time: Dec 04 10:51:09 2017 Quant Method : P:\HPCHEM1\FID_A\GASES\Method\FA112217 GASES.M Quant Title : QLast Update : Wed Nov 22 16:24:33 2017 Response via : Initial Calibration Integrator: ChemStation Volume Inj. : Signal Phase : RT-U Plot Signal Info : 30M x 0.32mm Min. RRF : 0.000 Min. Rel. Area : 50% Max. R.T. Dev 0.50min Max. RRF Dev : 25% Max. Rel. Area : 150% Compound AvgRF CCRF %Dev Area% Dev(Min) ------------------------------------------------------------------------- 1 Ethane 49.713 55.273 -11.2 107 -0.01 2 Ethylene 49.336 52.331 -6.1 105 -0.02 3 Methane 34.956 36.412 -4.2 104 0.00 Evaluate Continuing Calibration Report - Not Founds -------------------------------------------------------------------------- (#) = Out of Range SPCC's out = 0 CCC's out = 0 FA112217 GASES.M Tue Dec 26 15:56:32 2017