INSTRUMENT CONTROL PARAMETERS: MSVOA_V ----------------------------------------- D:\MassHunter\GCMS\1\methods\MSVOA_V.M Thu May 03 15:15:48 2018 Control Information ------- ----------- Sample Inlet : GC Injection Source : Manual Mass Spectrometer : Enabled No Sample Prep method has been assigned to this method. GC GC Summary Run Time 17.67 min Post Run Time 0 min Oven Temperature Setpoint On (Initial) 45 °C Hold Time 4.07 min Post Run 70 °C Program #1 Rate 15 °C/min #1 Value 240 °C #1 Hold Time 0.6 min Equilibration Time 0.25 min Max Temperature 260 °C Maximum Temperature Override Enabled Slow Fan Disabled Front SS Inlet He Mode Split Heater On 200 °C Pressure On 14.909 psi Total Flow On 61.7 mL/min Septum Purge Flow On 5 mL/min Gas Saver Off Split Ratio 80 :1 Split Flow 56 mL/min Thermal Aux 2 (MSD Transfer Line) Temperature Setpoint On (Initial) 250 °C Post Run 0 °C Column Column #1 Flow Setpoint On (Initial) 0.7 mL/min Post Run 0.57353 mL/min Agilent 121-1324E DB-624 0 °C—280 °C (280 °C): 20 m x 180 μm x 1 μm Column lock Unlocked In Front SS Inlet He Out MSD (Initial) 45 °C Pressure 14.909 psi Flow 0.7 mL/min Average Velocity 37.252 cm/sec Holdup Time 0.89482 min Column Outlet Pressure 0 psi Signals Signal #1: Test Plot Description Test Plot Details Save Off Data Rate 50 Hz Dual Injection Assignment Front Sample Signal #2: Test Plot Description Test Plot Details Save Off Data Rate 50 Hz Dual Injection Assignment Back Sample Signal #3: Test Plot Description Test Plot Details Save Off Data Rate 50 Hz Dual Injection Assignment Back Sample Signal #4: Test Plot Description Test Plot Details Save Off Data Rate 50 Hz Dual Injection Assignment Back Sample MS Information -- ----------- General Information ------- ----------- Acquisition Mode : Scan Solvent Delay (minutes) : 0.8 Tune file : D:\MassHunter\GCMS\1\5977\042418.u EM Setting mode Gain : 3.500000 Normal or Fast Scanning : Normal Scanning Trace Ion Detection : On Run Time (if MS only) : 10 minutes [Scan Parameters] Start Time : 0.8 Low Mass : 35 High Mass : 300 Threshold : 100 A/D Samples: : 4 [MSZones] MS Source : 250 C maximum 250 C MS Quad : 200 C maximum 200 C Timed Events ----- ------ Number Events= 0 END OF MS ACQUISTION PARAMETERS TUNE PARAMETERS for SN: US1635M037 --------------------------------- Trace Ion Detection is ON. 27.801 : EMISSION 70.007 : ENERGY 37.478 : REPELLER 86.592 : IONFOCUS 4.111 : ENTRANCE_LENS 1078.374 : EMVOLTS 1144.1 : Actual EMV 3.50 : GAIN FACTOR 2506.000 : AMUGAIN 139.000 : AMUOFFSET 2.000 : FILAMENT 1.000 : DCPOLARITY 8.289 : ENTLENSOFFSET 0.000 : Ion_Body 0.000 : EXTLENS 154.000 : MASSGAIN -24.000 : MASSOFFSET END OF TUNE PARAMETERS ---------------------- END OF INSTRUMENT CONTROL PARAMETERS ------------------------------------