Quantitation Report (QT/LSC Reviewed) Data Path : W:\HPCHEM1\MSVOA_V\DATA\VV052618\ Data File : VV006066.D Acq On : 26 May 2018 19:45 Operator : SY/MD Sample : J3090-17 100X Misc : 5.0 mL/MSVOA_V/WATER ALS Vial : 20 Sample Multiplier: 1 Quant Time: May 28 03:51:26 2018 Quant Method : W:\HPCHEM1\MSVOA_V\METHOD\SOMVLM052518WMA.M Quant Title : VOC Analysis QLast Update : Mon May 28 02:03:29 2018 Response via : Initial Calibration Internal Standards R.T. QIon Response Conc Units Dev(Min) -------------------------------------------------------------------------- 1) 1,4-Difluorobenzene 5.67 114 172870 50.00 ug/L 0.00 28) Chlorobenzene-d5 8.90 117 157616 50.00 ug/L 0.00 60) 1,4-Dichlorobenzene-d4 11.30 152 61216 50.00 ug/L 0.00 System Monitoring Compounds 4) Vinyl Chloride-d3 1.32 65 54825m 45.41 ug/L 0.00 Spiked Amount 50.000 Range 60 - 135 Recovery = 90.82% 7) Chloroethane-d5 1.55 69 7992 9.36 ug/L -0.02 Spiked Amount 50.000 Range 70 - 130 Recovery = 18.72%# 11) 1,1-Dichloroethene-d2 2.11 63 79014 35.16 ug/L -0.03 Spiked Amount 50.000 Range 60 - 125 Recovery = 70.32% 21) 2-Butanone-d5 3.96 46 54197 78.37 ug/L 0.01 Spiked Amount 100.000 Range 40 - 130 Recovery = 78.37% 24) Chloroform-d 4.40 84 104624 45.97 ug/L 0.00 Spiked Amount 50.000 Range 70 - 125 Recovery = 91.94% 26) 1,2-Dichloroethane-d4 5.09 65 73515 49.49 ug/L 0.00 Spiked Amount 50.000 Range 70 - 125 Recovery = 98.98% 32) Benzene-d6 5.10 84 208400 49.74 ug/L 0.00 Spiked Amount 50.000 Range 70 - 125 Recovery = 99.48% 36) 1,2-Dichloropropane-d6 6.12 67 67009 50.36 ug/L 0.00 Spiked Amount 50.000 Range 70 - 120 Recovery = 100.72% 41) Toluene-d8 7.36 98 182533 47.09 ug/L 0.00 Spiked Amount 50.000 Range 80 - 120 Recovery = 94.18% 43) trans-1,3-Dichloropropene- 7.67 79 27022 44.09 ug/L 0.00 Spiked Amount 50.000 Range 60 - 125 Recovery = 88.18% 47) 2-Hexanone-d5 8.15 63 49826 95.27 ug/L 0.01 Spiked Amount 100.000 Range 45 - 130 Recovery = 95.27% 57) 1,1,2,2-Tetrachloroethane- 10.27 84 89133 48.71 ug/L 0.00 Spiked Amount 50.000 Range 65 - 120 Recovery = 97.42% 64) 1,2-Dichlorobenzene-d4 11.68 152 66099 54.29 ug/L 0.00 Spiked Amount 50.000 Range 80 - 120 Recovery = 108.58% Target Compounds Qvalue 34) Trichloroethene 5.96 95 49801 39.18 ug/L 98 -------------------------------------------------------------------------- (#) = qualifier out of range (m) = manual integration (+) = signals summed