INSTRUMENT CONTROL PARAMETERS: MSVOA_W ----------------------------------------- D:\MassHunter\GCMS\1\methods\MSVOA_W.M Fri May 25 10:11:31 2018 Control Information ------- ----------- Sample Inlet : GC Injection Source : External Device Mass Spectrometer : Enabled No Sample Prep method has been assigned to this method. GC GC Summary Run Time 16.818 min Post Run Time 0 min Oven Equilibration Time 1 min Max Temperature 450 °C Maximum Temperature Override Enabled Slow Fan Disabled Temperature Setpoint On (Initial) 40 °C Hold Time 5 min Post Run 70 °C Program #1 Rate 11 °C/min #1 Value 60 °C #1 Hold Time 0 min #2 Rate 20 °C/min #2 Value 220 °C #2 Hold Time 2 min Front SS Inlet He Mode Split Heater On 200 °C Pressure On 3.8551 psi Total Flow On 52.7 mL/min Septum Purge Flow On 3 mL/min Gas Saver Off Split Ratio 70 :1 Split Flow 49 mL/min Thermal Aux 2 (MSD Transfer Line) Temperature Setpoint On (Initial) 250 °C Post Run 0 °C Column Column Outlet Pressure 0 psi Column #1 HP-5MS 5% Phenyl Methyl Silox Temperature Range 0 °C—325 °C (325 °C) Dimensions 30 m x 250 μm x 1.4 μm Column lock Unlocked In Front SS Inlet He Out MSD (Initial) 40 °C Pressure 3.8551 psi Flow 0.7 mL/min Average Velocity 30.339 cm/sec Holdup Time 1.6481 min Flow Setpoint On (Initial) 0.7 mL/min Post Run 1 mL/min Signals Signal #1: Test Plot Description Test Plot Details Save Off Data Rate 50 Hz Signal #2: Test Plot Description Test Plot Details Save Off Data Rate 50 Hz Signal #3: Test Plot Description Test Plot Details Save Off Data Rate 50 Hz Signal #4: Test Plot Description Test Plot Details Save Off Data Rate 50 Hz MS Information -- ----------- General Information ------- ----------- Acquisition Mode : Scan Solvent Delay (minutes) : 1.5 Tune file : D:\MassHunter\GCMS\1\5977\040918.u EM Setting mode Gain : 2.500000 Normal or Fast Scanning : Normal Scanning Trace Ion Detection : Off Run Time (if MS only) : 650 minutes [Scan Parameters] Start Time : 1.5 Low Mass : 35 High Mass : 300 Threshold : 50 A/D Samples: : 8 [MSZones] MS Source : 250 C maximum 250 C MS Quad : 200 C maximum 200 C Timed Events ----- ------ Number Events= 0 END OF MS ACQUISTION PARAMETERS TUNE PARAMETERS for SN: US1735M030 --------------------------------- Trace Ion Detection is OFF. 18.201 : EMISSION 70.007 : ENERGY 30.019 : REPELLER 114.289 : IONFOCUS 8.907 : ENTRANCE_LENS 982.604 : EMVOLTS 999.5 : Actual EMV 2.50 : GAIN FACTOR 2122.000 : AMUGAIN 138.438 : AMUOFFSET 2.000 : FILAMENT 1.000 : DCPOLARITY 14.443 : ENTLENSOFFSET 0.000 : Ion_Body 0.000 : EXTLENS -65.000 : MASSGAIN -19.000 : MASSOFFSET END OF TUNE PARAMETERS ---------------------- END OF INSTRUMENT CONTROL PARAMETERS ------------------------------------