Quantitation Report (QT/LSC Reviewed) Data Path : W:\HPCHEM1\MSVOA_X\DATA\VX050818\ Data File : VX001505.D Acq On : 08 May 2018 11:10 Operator : JC/MD Sample : VX0508WBL01 Misc : 5.0mL/MSVOA_X/WATER ALS Vial : 3 Sample Multiplier: 1 Quant Time: May 08 13:55:48 2018 Quant Method : W:\HPCHEM1\MSVOA_X\METHOD\82X050518W.M Quant Title : SW846 8260 QLast Update : Sun May 06 07:59:25 2018 Response via : Initial Calibration Internal Standards R.T. QIon Response Conc Units Dev(Min) -------------------------------------------------------------------------- 1) Pentafluorobenzene 5.67 168 288752 50.00 ug/l 0.00 34) 1,4-Difluorobenzene 6.87 114 389214 50.00 ug/l 0.00 63) Chlorobenzene-d5 10.12 117 335286 50.00 ug/l 0.00 72) 1,4-Dichlorobenzene-d4 12.09 152 181113 50.00 ug/l 0.00 System Monitoring Compounds 33) 1,2-Dichloroethane-d4 6.07 65 231870 55.87 ug/l 0.00 Spiked Amount 50.000 Recovery = 111.74% 35) Dibromofluoromethane 5.51 113 181456 52.41 ug/l 0.00 Spiked Amount 50.000 Recovery = 104.82% 50) Toluene-d8 8.72 98 666329 52.73 ug/l 0.00 Spiked Amount 50.000 Recovery = 105.46% 62) 4-Bromofluorobenzene 11.14 95 216775 44.69 ug/l 0.00 Spiked Amount 50.000 Recovery = 89.38% Target Compounds Qvalue -------------------------------------------------------------------------- (#) = qualifier out of range (m) = manual integration (+) = signals summed