Quantitation Report (QT/LSC Reviewed) Data Path : W:\HPCHEM1\MSVOA_X\DATA\VX051118\ Data File : VX001637.D Acq On : 11 May 2018 13:53 Operator : JC/MD Sample : VX0511WBL02 Misc : 5.0mL/MSVOA_X/WATER ALS Vial : 3 Sample Multiplier: 1 Quant Time: May 12 05:58:10 2018 Quant Method : W:\HPCHEM1\MSVOA_X\METHOD\82X050518W.M Quant Title : SW846 8260 QLast Update : Sun May 06 07:59:25 2018 Response via : Initial Calibration Internal Standards R.T. QIon Response Conc Units Dev(Min) -------------------------------------------------------------------------- 1) Pentafluorobenzene 5.67 168 244204 50.00 ug/l 0.00 34) 1,4-Difluorobenzene 6.87 114 326186 50.00 ug/l 0.00 63) Chlorobenzene-d5 10.12 117 274327 50.00 ug/l 0.00 72) 1,4-Dichlorobenzene-d4 12.09 152 137878 50.00 ug/l 0.00 System Monitoring Compounds 33) 1,2-Dichloroethane-d4 6.07 65 193969 55.26 ug/l 0.00 Spiked Amount 50.000 Recovery = 110.52% 35) Dibromofluoromethane 5.51 113 153577 52.93 ug/l 0.00 Spiked Amount 50.000 Recovery = 105.86% 50) Toluene-d8 8.72 98 546561 51.61 ug/l 0.00 Spiked Amount 50.000 Recovery = 103.22% 62) 4-Bromofluorobenzene 11.14 95 171817 42.26 ug/l 0.00 Spiked Amount 50.000 Recovery = 84.52% Target Compounds Qvalue -------------------------------------------------------------------------- (#) = qualifier out of range (m) = manual integration (+) = signals summed