Quantitation Report (QT/LSC Reviewed) Data Path : W:\HPCHEM1\MSVOA_X\DATA\VX051718\ Data File : VX001785.D Acq On : 17 May 2018 12:52 Operator : JC/MD Sample : J2796-04 Misc : 5.0mL/MSVOA_X/WATER ALS Vial : 5 Sample Multiplier: 1 Quant Time: May 18 04:27:43 2018 Quant Method : W:\HPCHEM1\MSVOA_X\METHOD\82X051518W.M Quant Title : SW846 8260 QLast Update : Tue May 15 01:36:22 2018 Response via : Initial Calibration Internal Standards R.T. QIon Response Conc Units Dev(Min) -------------------------------------------------------------------------- 1) Pentafluorobenzene 5.67 168 302010 50.00 ug/l 0.00 34) 1,4-Difluorobenzene 6.87 114 410140 50.00 ug/l 0.00 63) Chlorobenzene-d5 10.12 117 328413 50.00 ug/l 0.00 72) 1,4-Dichlorobenzene-d4 12.09 152 174561 50.00 ug/l 0.00 System Monitoring Compounds 33) 1,2-Dichloroethane-d4 6.07 65 225965 55.82 ug/l 0.00 Spiked Amount 50.000 Recovery = 111.64% 35) Dibromofluoromethane 5.51 113 192186 55.53 ug/l 0.00 Spiked Amount 50.000 Recovery = 111.06% 50) Toluene-d8 8.72 98 682594 55.06 ug/l 0.00 Spiked Amount 50.000 Recovery = 110.12% 62) 4-Bromofluorobenzene 11.14 95 196959 44.13 ug/l 0.00 Spiked Amount 50.000 Recovery = 88.26% Target Compounds Qvalue 16) Acetone 2.45 43 10833 6.81 ug/l 95 27) cis-1,2-Dichloroethene 4.60 96 8819 2.36 ug/l 95 64) Tetrachloroethene 9.34 164 1261 0.29 ug/l # 89 -------------------------------------------------------------------------- (#) = qualifier out of range (m) = manual integration (+) = signals summed