Quantitation Report (QT Reviewed) Data Path : W:\HPCHEM1\MSVOA_X\DATA\VX051718\ Data File : VX001799.D Acq On : 17 May 2018 19:07 Operator : JC/MD Sample : J2853-07DL 10X Misc : 5.0mL/MSVOA_X/WATER ALS Vial : 19 Sample Multiplier: 1 Quant Time: May 18 10:30:38 2018 Quant Method : W:\HPCHEM1\MSVOA_X\METHOD\82X051518W.M Quant Title : SW846 8260 QLast Update : Tue May 15 01:36:22 2018 Response via : Initial Calibration Internal Standards R.T. QIon Response Conc Units Dev(Min) -------------------------------------------------------------------------- 1) Pentafluorobenzene 5.67 168 261510 50.00 ug/l 0.00 34) 1,4-Difluorobenzene 6.87 114 342510 50.00 ug/l 0.00 63) Chlorobenzene-d5 10.12 117 270865 50.00 ug/l 0.00 72) 1,4-Dichlorobenzene-d4 12.08 152 133560 50.00 ug/l 0.00 System Monitoring Compounds 33) 1,2-Dichloroethane-d4 6.07 65 155743 44.43 ug/l 0.00 Spiked Amount 50.000 Recovery = 88.86% 35) Dibromofluoromethane 5.51 113 132452 45.82 ug/l 0.00 Spiked Amount 50.000 Recovery = 91.64% 50) Toluene-d8 8.72 98 451451 43.61 ug/l 0.00 Spiked Amount 50.000 Recovery = 87.22% 62) 4-Bromofluorobenzene 11.15 95 129305 34.69 ug/l 0.00 Spiked Amount 50.000 Recovery = 69.38% Target Compounds Qvalue 4) Vinyl Chloride 1.40 62 13629 4.45 ug/l 99 27) cis-1,2-Dichloroethene 4.60 96 228762 70.60 ug/l 90 -------------------------------------------------------------------------- (#) = qualifier out of range (m) = manual integration (+) = signals summed